Bidspirit auction | ARLYTE IONOGRAM APPARATUS K, NA,
ARLYTE IONOGRAM APPARATUS K, NA, CL, CA, PH ANALYZER - This device measures Sodium, Potassium, Chloride, Calcium, and pH using the manometric method. Sample type to use: * Whole blood, serum and/or plasma (100 µl - 150 µl) * Pre-diluted urine (200 µl) * Data storage: Up to 1,000 samples with results every 24 hours Detection range: * whole blood, serum and plasma * K+: 0.5 - 15.0 mmol/L * Na+: 20 - 200 mmol/L * Cl-: 20 - 200 mmol/L * Ca²⁺: 0.1 - 5.0 mmol/L * pH: 4.0 - 9.0 pH Calibration of an ISE 6000 Ionogram 2-point calibration: Cal A: * K+: 4.0 mmol/L * Na+: 140.0 mmol/L * Cl-: 100.0 mmol/L * Ca²⁺: 1.0 mmol/L * pH: 7.4 Cal B: * K+: 8.0 mmol/L * Na+: 110.0 mmol/L * Cl-: 70.0 mmol/L * Ca²⁺: 2.0 mmol/L * pH: 7.8 * User interface: wide LCD screen * Thermal printer: 32 characters * Ambient temperature: 10° - 30° C * Humidity: < 70% * Serial port: RS 232 * Voltage:220 V +/- 10% or 110 V +/- 10%; 50Hz +/-1Hz * Power consumption: < 60 W * Dimensions: 50cm x 27cm x 36cm * Weight: 7.2 kg Precision : * K+: +/- 0.2 mmol/L * Na+: +/- 2.0 mmol/L * Cl-: +/- 2.5 mmol/L * Ca²⁺: +/- 0.1 mmol/L * pH: +/- 0.1 Reproducibility of an ionogram (whole blood, serum or plasma) * K+: < 1.5% * Na+: < 1.5% * Cl-: < 1.5% * Ca²⁺: < 3.0% * pH: < 1.0%